The voltage difference comes from the probes and their impedance.In HF it is necessary to use suitable probes / La differenza di tensione deriva dalle sonde e dalla loro impedenza. In HF è necessario utilizzare sonde adeguate
Late, but better than never. You should put the 50ohm on, the losses on cables and reflections are unpredictable. I was actually looking for a review on the 1062x Plus :)
The difference in amplitude could be because of reflection. Were the outputs of the generators set to HiZ or 50Ohm? The oscilloscope should also match these impedance. Are the cables of the same length? thanks for the comparison.
exactly the comparison I was looking for.Thanks / esattamente il confronto che stavo cercando. Grazie
The voltage difference comes from the probes and their impedance.In HF it is necessary to use suitable probes / La differenza di tensione deriva dalle sonde e dalla loro impedenza. In HF è necessario utilizzare sonde adeguate
Late, but better than never. You should put the 50ohm on, the losses on cables and reflections are unpredictable.
I was actually looking for a review on the 1062x Plus :)
The difference in amplitude could be because of reflection. Were the outputs of the generators set to HiZ or 50Ohm? The oscilloscope should also match these impedance. Are the cables of the same length? thanks for the comparison.
+HerrSattarkosej good point, both gen output HiZ and scope input 1MOhm
Probabilmente è dovuto ai filtri passabasso in uscita diversi tra i due modelli. A 40 Mhz siamo comunque al limite di banda per il SDG2042X