  • How to conduct calibration of a #pH #meter?
    (Mettler-Toledo S220)
    Good pH Practice:
    1. Always calibrate your sensor before the start of measurement
    2. Do no resue buffer solutions
    3. Discard opened solution after 6 months
    4. Allow sensors to equilibrate with the environment before each measurement
    5. Wait for reading stabilized icon to appear before recording the measurement
    6. Recharge or replace sensor when
    - Calibration slope falls within 85%-89%
    - Offset value more than 35mV
    - Slow response time during pH measurement and calibration
    7. Ony store sensor in 3.0M KCl
    8. Do not wipe sensor in harsh motion
    9. NEVER store sensor in deionized water
    🙏🏼Thank you so much, do give us a 'LIKE' and Subscribe🙏🏼!
    🤗hope to see you again🤗!
    🌟Wish you all the best in life.🌻!
    Here in this video, we have an expert who guide you the good way with the IPOV (Instructor's point of view) technique.
    As always, be safe, and learn well.
    Click to access to more great videos
    • Playlist
    Subscribe to receive videos on new instruments, machines in science, chemistry.
    If you would like to have more chemistry fun, and learn about cool science, subscribe to this channel to view the upcoming videos.
    Thanks for watching and never give up in whatever you do !
    REMEMBER: “The beginning is always the hardest”
    LIGHT-HEARTED CHEMISTRY LECTURE SERIES - FUN MAN FLIPPED CLASSROOM
    Carboxylic Acid Derivative
    • Video
    Alkene
    • Video
    Inductive - Resonance Effect
    • Video
    VIDEOS ON CHEMISTRY TECHNIQUES IN THE LAB
    UV Spectroscopy
    • UV Vis spectroscopy 🌟
    Thin Layer Chromatography
    • 【4K】 --- Thin Layer Ch...
    Flash Column Chromatography
    • 【4K】-- Column Chromato...
    NMR Nuclear Magnetic Resonance
    • NMR Spectrum Processin...
    Liquid-liquid extraction
    • 🌟 -- Liquid-Liquid Ext...
    Folding Fluted Filter paper
    • 🌟 Filter Paper Chemist...
    ChemDraw Pro 15.0 Tutorial
    • ChemDraw Tutorial Basi...
    Schlenk Line
    • Schlenk Line Technique...
    Glove Box
    • 🌟 -- Glove Box - opera...
  • วิทยาศาสตร์และเทคโนโลยี

ความคิดเห็น • 89

  • @d.deepak
    @d.deepak 4 ปีที่แล้ว +6

    great explanation, got my pH meter calibrated with your help. Thanks

  • @xavik
    @xavik 8 ปีที่แล้ว +1

    Thanks! I already missed these videos.

    • @chemfunman
      @chemfunman  8 ปีที่แล้ว

      you are most welcome :')

  • @yohanseol6296
    @yohanseol6296 6 ปีที่แล้ว +1

    thank you very much!!^^

  • @longho2097
    @longho2097 4 ปีที่แล้ว +1

    when i select Menu structure of pH/ion, the display only show two mode 1 Sensor ID/ IM and 2 Calibration Setting. I choose mode 2 but the machine don't show mode Buffer Group/ Standard. On the other hand, at the main display, the machine in Router mode, i can't calibration the group buffer 4, 7, 9.21 because this machine mesure not exactly. Can you help me the method to this machine return as originally, when i press menu structure of pH/ion have 7 mode. Thank you!!

  • @chemfunman
    @chemfunman  4 ปีที่แล้ว +2

    SUBSCRIBE NOW to this channel now to receive more great chem videos!
    Connect with me on
    Facebook | facebook.com/chemfunman/
    Twitter @chemfunman (twitter.com/chemfunman)
    Stay safe!
    “Ever tried. Ever failed. No Matter. Try again. Fail again. Fail better.”
    - Samuel Beckett

  • @mightyman6459
    @mightyman6459 6 ปีที่แล้ว +3

    Very useful video, thanks teacher!

  • @alyaae.h3533
    @alyaae.h3533 7 ปีที่แล้ว +1

    thank you...

  • @wafakallaa5152
    @wafakallaa5152 6 ปีที่แล้ว +1

    thanks a lot

  • @JashanjotKaur
    @JashanjotKaur 3 ปีที่แล้ว +1

    Very well explained. Thanks.

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว +1

      thank you very much! please subscribe and share this video with your friends!

  • @pavadaiarumugam9723
    @pavadaiarumugam9723 3 ปีที่แล้ว +2

    Very clear explanation. Thank you so much sir

  • @Sp-fe1eo
    @Sp-fe1eo 2 ปีที่แล้ว +1

    thanks for the video...it's really useful

    • @chemfunman
      @chemfunman  2 ปีที่แล้ว

      You are most welcome

  • @parimisantu5148
    @parimisantu5148 7 ปีที่แล้ว +1

    Nice

  • @guncix21
    @guncix21 3 ปีที่แล้ว +1

    When you take the KCl conditioning vial of the pH probe, you have to hold the clear vial part and twist the white part anti-clockwise, therefore the vile will be gently released. When you finish analysis put it back on and twist it clockwise direction to seal it.

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      thank you very much!

  • @ashrafgafar
    @ashrafgafar 3 ปีที่แล้ว +2

    This was brilliant

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว +1

      Thank you so much for your support. God bless and subscribe, share this love or learning with others. Cheers.

  • @medosou
    @medosou 5 ปีที่แล้ว +2

    Very well demonstrated

  • @jhogan1960
    @jhogan1960 5 ปีที่แล้ว +3

    Interesti9ng how they do things in different countries. I always learn something New. Hello from the USA!

    • @chemfunman
      @chemfunman  5 ปีที่แล้ว

      appreciate your comment! hello from Singapore, James!

  • @afzalshahbaz89
    @afzalshahbaz89 5 ปีที่แล้ว +5

    at 2:05 you don't have to pull out the probe from the 3M KCL soaking tube. Simply rotate the upper flexible screw of the soaking tube and the probe will come out itself without any forced rotation of the soaking tube from the bottom. Else it may damage the probe bulb in process.

    • @chemfunman
      @chemfunman  4 ปีที่แล้ว

      afzal shahbaz that’s right!

  • @RajendraSingh-ps6of
    @RajendraSingh-ps6of 5 ปีที่แล้ว +1

    How to plot pH graph with different buffer solution

  • @thembelangxokile1571
    @thembelangxokile1571 7 ปีที่แล้ว +3

    it would be helpful though if you told us why we do certain things like taking steps to remove the bubbles

    • @chemfunman
      @chemfunman  7 ปีที่แล้ว +1

      would you like to chip in , appreciate if all do out part ! ^^

  • @yasiralshebib6626
    @yasiralshebib6626 3 ปีที่แล้ว +2

    Hi Dr Fun
    thanks for the education video.
    I'd like to ask you about: (#good safety habit), is it just a hashtag on twitter or there is a youtube video channel dedicated to safety habits videos?
    thanks in advance

  • @matchmedicines8209
    @matchmedicines8209 ปีที่แล้ว +1

    thanks a lot, it was amazing.
    just a question, how I can prepare the buffers (4,9, and 7)?

    • @chemfunman
      @chemfunman  ปีที่แล้ว

      we bought them from the store.

  • @amenananias2134
    @amenananias2134 ปีที่แล้ว +1

    A good video

    • @chemfunman
      @chemfunman  ปีที่แล้ว

      Glad you enjoyed it

  • @shanrao5388
    @shanrao5388 3 ปีที่แล้ว +1

    What type of error gave the PH meter to find out PH..?

  • @lydiaamelia199
    @lydiaamelia199 3 ปีที่แล้ว +2

    Great and clear explanation, i love that, but I have a question, what we have to do if the pH meter measure wrongly or inaccurate when calibrating the buffer solution, like when the actual pH value of the buffer is 4.01 but the pH meter show the value of 4.51 then what should we do?

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      plot a calibration curve, and adjust accordingly

  • @MaestroAssistant
    @MaestroAssistant 4 ปีที่แล้ว +1

    I think you can rotate white plastic top to unscrew it from the KCl container

    • @chemfunman
      @chemfunman  4 ปีที่แล้ว

      Daniel Arroyo you are right!

  • @suppachaikrajangsri2196
    @suppachaikrajangsri2196 3 ปีที่แล้ว +1

    Hello, I have a problem with Cal2, it show wrong buffer, what could be wrong?

    • @chemfunman
      @chemfunman  2 ปีที่แล้ว

      the comments show "wrong buffer"? check if the buffer is new. because it is prone to pH change when used for a long time.

  • @jawadnadeem2920
    @jawadnadeem2920 5 ปีที่แล้ว +1

    what if my first two calibrations go well and the third one shows 10.01 instead of 9.21?? what could go wrong?? possibly it does not measure properly above 7???

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      did you try again?

  • @jaswanthchowdary17-9-21
    @jaswanthchowdary17-9-21 6 ปีที่แล้ว +1

    Sir I want slope calculation on 3 buffers calibration ND u miss the slope value in this video...

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      slope is the measure of the linearity of the calibration - is it accurate? can we believe the readings? with what type of precisions?

  • @muntahafazal237
    @muntahafazal237 5 ปีที่แล้ว +1

    Wow..

    • @chemfunman
      @chemfunman  5 ปีที่แล้ว

      Muntaha Fazal that’s an old game

    • @chemfunman
      @chemfunman  5 ปีที่แล้ว

      Muntaha Fazal legendary

  • @bouallalahalima6459
    @bouallalahalima6459 2 ปีที่แล้ว +1

    Thank's but I have a problem when l calibration read only one calibration 1 don't read calibration 2

  • @uartim
    @uartim ปีที่แล้ว +1

    Do you need a buffer with higher pH value for solution to be measured at a higher pH ?

    • @chemfunman
      @chemfunman  ปีที่แล้ว

      need it for calibration.

  • @jomarcamu7057
    @jomarcamu7057 5 ปีที่แล้ว +5

    Seriously do you think you did it right? You calibrated your instrument by standard buffer 4.01, followed by 9.21 and finally by 7.00. And after calibration you pressed exit button without even pressing END button hence not saving your calibration into your instrument. Consequently no slope can be displayed by your instrument since you didn't even ended your calibration and saved it. Correct me if I am wrong if you disagree.

    • @hafizusamma1497
      @hafizusamma1497 5 ปีที่แล้ว

      Good Question. Would you please, refer me some of best companies to purchase a Ph meter? One reliable, robust too for industrial use. Thanks! (hafizusamma@mpcl.com.pk)

    • @ihartu5549
      @ihartu5549 ปีที่แล้ว

      he calibrated the buffer ph7 first 😏 lmao don't comment when you can't comprehend.. and one more thing, not all ph meter are the same...

  • @alial-absawi8880
    @alial-absawi8880 5 หลายเดือนก่อน +1

    Hello
    I hope you are doing well
    May I ask that why u didn't press End after that Save
    Also I have problem with my ph meter which one of the puffer always can not be celebrated

    • @chemfunman
      @chemfunman  5 หลายเดือนก่อน +1

      I don’t understand your last sentence

    • @alial-absawi8880
      @alial-absawi8880 5 หลายเดือนก่อน

      Sorry I meant calibrated @4&7ok but when I come to 10 display showed out of the range

  • @jerilyncruz5069
    @jerilyncruz5069 4 ปีที่แล้ว

    Do you only need to calibrate once or every use? Or just after every session?

  • @ZeMagnifier
    @ZeMagnifier 8 หลายเดือนก่อน +1

    Won't we risk contaminating the buffer solutions if we were to directly dip the proble into the buffer solution storage?

    • @chemfunman
      @chemfunman  8 หลายเดือนก่อน

      can wipe quickly

    • @ZeMagnifier
      @ZeMagnifier 8 หลายเดือนก่อน

      @@chemfunman I don't mean the probe itself, I mean the buffer solution. Is it still as accurate as the original measurements before callibrating the meter?

  • @agustinnovalinda6468
    @agustinnovalinda6468 2 ปีที่แล้ว

    Can u explain hau to change temperature setting from fareinheit to celcius on display, please share

  • @hamzaelfadili-7875
    @hamzaelfadili-7875 6 ปีที่แล้ว +1

    please what's the role of Kcl

    • @chemfunman
      @chemfunman  6 ปีที่แล้ว

      prevent leaching of water via osmosis. if this happens, the KCl crystallize. forming an ionic salt. immobilized ions

    • @chemfunman
      @chemfunman  6 ปีที่แล้ว

      prevent leaching of water via osmosis. if this happens, the KCl crystallize. forming an ionic salt. immobilized ions

    • @chemfunman
      @chemfunman  6 ปีที่แล้ว

      prevent leaching of water via osmosis. if this happens, the KCl crystallize. forming an ionic salt. immobilized ions

  • @cosmeticreviews9
    @cosmeticreviews9 4 ปีที่แล้ว +1

    Please tell what does offset and slope mean in pH meter?

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      offset is the variance - inherent error by any instrument

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      slope is the measure of the linearity of the calibration - is it accurate? can we believe the readings? with what type of precisions?

  • @saikorada8342
    @saikorada8342 11 หลายเดือนก่อน +1

    Sir telugu mediam sir please

  • @SVERMA941
    @SVERMA941 5 ปีที่แล้ว +1

    What is purpose of slop and offset limit?

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว +1

      slope is the measure of the linearity of the calibration - is it accurate? can we believe the readings? with what type of precisions?

    • @SVERMA941
      @SVERMA941 3 ปีที่แล้ว +1

      @@chemfunman After calibration u can verify by buffer

  • @mimikiongmeimei6659
    @mimikiongmeimei6659 5 ปีที่แล้ว +1

    why use pH 7 first?

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      neutrality is always unbiased. that is the best reference.

  • @myassarkhiami6442
    @myassarkhiami6442 3 ปีที่แล้ว +1

    I need this machine

    • @chemfunman
      @chemfunman  3 ปีที่แล้ว

      This machine is BEAST.

  • @Joshua-dc1bs
    @Joshua-dc1bs 5 ปีที่แล้ว +2

    My favourite part of my job is staring at the screen as it... slowly... calibrates. Weighing scales are even worse :/

  • @shi__532
    @shi__532 4 ปีที่แล้ว +2

    haha Slam dunk enthusiast??

    • @chemfunman
      @chemfunman  4 ปีที่แล้ว

      Syifaq Azlan not bad you recognise!

    • @chemfunman
      @chemfunman  4 ปีที่แล้ว +1

      Syifaq Azlan very good series

    • @shi__532
      @shi__532 4 ปีที่แล้ว +1

      @@chemfunman childhood anime