How to calibrate your pH electrode

แชร์
ฝัง
  • เผยแพร่เมื่อ 2 ต.ค. 2024

ความคิดเห็น • 36

  • @akakisimonishvili7471
    @akakisimonishvili7471 6 หลายเดือนก่อน +1

    Hello, I have a new HACH PHC101 electrode, but I can not calibrate, it says (standard not recognized. Rinse the electrode and repeat). Can you help me and tell me the reason?

    • @MetrohmTV
      @MetrohmTV  5 หลายเดือนก่อน

      Hi
      Thank you for your comment.
      Could you please check if another electrode can be calibrated? This way we would know if the problem lies with the instrument or the sensor.
      In general, if you require instrument support, please contact your local Metrohm organization or exclusive distributor.
      BR
      Michel, Metrohm International Headquarters

  • @changyongkang252
    @changyongkang252 ปีที่แล้ว +1

    Why the calibrate during first buffer taking much, sometimes wait around 45 minute still cannot proceed to next buffer

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว +1

      Hi
      Thank you for your comment. Typically, a buffer measurement should not take more than a few seconds or a few minutes at most. A measuring time of 45 minutes is a clear indicator that something is not right, the most obvious candidate being a defective electrode. Please check whether your electrode needs to be replaced or contact your local Metrohm organization if you need technical support:
      www.metrohm.com/en.html#country-selection
      Best regards
      Metrohm International Headquarters

  • @RobIncFilmsWeddings
    @RobIncFilmsWeddings ปีที่แล้ว +3

    Thank you for running over the basics of pH calibration and a reminder to calibrate more often to ensure consistent sample accuracy!

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      Glad it was helpful!

    • @RobIncFilmsWeddings
      @RobIncFilmsWeddings ปีที่แล้ว

      @@MetrohmTV Is it still advisable to use pH 9 buffer for calibration bracketing samples with expected higher pH values, even though alkaline buffers have a high CO2 uptake?

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      For precise pH measurement results, we always recommend a multi-point calibration. The pH of the sample should always be within the pH range of the buffers that are used for calibration to minimize errors. So, if you are expecting higher pH values (e.g., pH 8) the use of pH buffer 9 is advisable.
      To avoid a carbon dioxide uptake of buffers with higher pH values, always use fresh buffer solutions. Our buffer solutions in single use sachets are quite suitable for this.
      If you have more questions on pH measurement and calibration please have a look at our current monograph “pH-measurement - everything from A-Z including best practices and examples”:
      www.metrohm.com/en/products/8/1085/81085081.html

  • @tnktxn640
    @tnktxn640 3 หลายเดือนก่อน

    Why do I have to calibrate more often when performed at high temperatures? If I do it at 28 degrees Celsius, is calibration necessary every day?

    • @MetrohmTV
      @MetrohmTV  3 หลายเดือนก่อน

      We would recommend calibrating the sensor daily and at the temperature of the samples for exact results. It depends a bit on the accuracy that is required.

  • @ericksanchezperez829
    @ericksanchezperez829 5 หลายเดือนก่อน

    What is the problem when the slope is Less than 97%?
    And pH(0) is higger than 7.10?

    • @MetrohmTV
      @MetrohmTV  5 หลายเดือนก่อน

      Hi Erick
      There could be various reasons for this. You can find additional information on this topic in our Application Bulletin AB-188:
      www.metrohm.com/en/applications/ab-application-bulletins/ab-188.html
      If you need to troubleshoot your instrument or sensor, please get in touch with your local Metrohm subsidiary or exclusive distributor:
      www.metrohm.com/en.html#country-selection
      BR
      Michel, Metrohm International Headquarters

  • @gidissadeksissa2219
    @gidissadeksissa2219 ปีที่แล้ว +1

    thank you for the detailed explanation
    Addition of NaOH or HCl to adjust the PH of a solution does not affect the initial concentration of a solution?

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      Hi gidissadeksissa2219
      Do you mean the initial concentration of an analyte? The initial concentration of an analyte is not affected if the analyte is sodium, hydroxide, hydronium, or chloride. Does this answer your question, or did we misunderstand it?
      Best regards
      Metrohm International Headquarters

  • @refar1624
    @refar1624 7 หลายเดือนก่อน +1

    Muy buen video y lo explica muy bien.

  • @BenoitPlomb-u8l
    @BenoitPlomb-u8l หลายเดือนก่อน

    What maintenance intervals (calibration, electrolyte replacement) would you recommend for a pH electrode that is used very rarely, e.g. once a year?

    • @MetrohmTV
      @MetrohmTV  หลายเดือนก่อน +1

      Hi
      This very much depends on the accuracy of the result you want to achieve. For example, if the result is only a rough guide, it is fine to perform a calibration or replace the electrolyte before the pH measurement. However, if you want the result to be as accurate as possible, maintenance is more important and should be carried out at least monthly or even weekly.
      In general, frequent maintenance of the electrode is recommended for warranty reasons and also with regard to the lifetime of an electrode.

    • @BenoitPlomb-u8l
      @BenoitPlomb-u8l หลายเดือนก่อน

      @@MetrohmTV Thanks a lot for your answer !

    • @MetrohmTV
      @MetrohmTV  หลายเดือนก่อน

      @@BenoitPlomb-u8l You're welcome. We hope this helps!

  • @kirankumargarud2800
    @kirankumargarud2800 ปีที่แล้ว

    Is it necessary to maintain the buffer temperature externally at exact 25 degree celcious during calibration and sample pH measurment as well

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      Since temperature has an impact on your pH measurement (Nernst equation), calibration and pH measurement should be performed at the same temperature, e.g., at 25 °C. For further information please read our blog post “How temperature influences the pH value”:
      www.metrohm.com/en/discover/blog/2023/how-temperature-influences-the-ph-value.html

  • @samueltinku250
    @samueltinku250 11 หลายเดือนก่อน

    Great video 😊! Why only 4,7,9 buffers used for calibration!
    What is the maximum limit can dip the ph electrode or need to only dip the end of the electrode

    • @MetrohmTV
      @MetrohmTV  11 หลายเดือนก่อน

      Hi Samuel
      Thank you for your comment. You always have to calibrate with buffers which lie within your measuring range. With buffers pH 4, 7, and 9, you’re already covering quite a broad pH range. Should you need to cover an even broader range, you can use pH buffers from other manufacturers.
      As for the electrode: The electrode has to be dipped in in such a way that the diaphragm is fully covered by your measuring solution. I.e., the glass membrane and diaphragm have to be immersed but the electrolyte filling hole has to stay above the sample.
      Best regards
      Michel, Metrohm International Headquarters

    • @samueltinku250
      @samueltinku250 11 หลายเดือนก่อน

      @@MetrohmTV 😀 thanks for your reply 😊

  • @srilatasampath9883
    @srilatasampath9883 ปีที่แล้ว

    Where do we store the buffer solution if we are not making fresh everyday

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      Hi
      Buffer sachets are for single use only. Metrohm buffers in the bottle last up to 3 months after opening and are easy to store on shelves. After a calibration, the buffer should be discarded and should not be reused.
      You can find more information in this blog post:
      www.metrohm.com/en/discover/blog/20-21/faq--all-about-ph-calibration.html

  • @farooqarshad5962
    @farooqarshad5962 ปีที่แล้ว

    What is ideal slope value after pH calibration.

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      The electrode slope is given by the Nernst equation and is 0.059 V per pH 1 at 25 °C. These are ideal values from which new Metrohm pH electrodes only differ slightly. The electrode zero point is ± 0.015 V. But please be aware that the electrode zero point and the electrode slope may change as a result of the aging of the glass membrane or changes (e.g., contamination) on the diaphragm.
      For more information on this topic, please read our blog “FAQ: All about pH calibration”:
      www.metrohm.com/en/discover/blog/20-21/faq--all-about-ph-calibration.html
      And our our application bulletin AB-188 “pH measurement technique”:
      www.metrohm.com/en/applications/ab-application-bulletins/ab-188.html

  • @iviaggidiAntony
    @iviaggidiAntony ปีที่แล้ว

    I always use three buffer : pH 4 - 7 - 9

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      That's exactly the right way to do it!😊

  • @prashantpatil5766
    @prashantpatil5766 ปีที่แล้ว

    Thanks for the detailed calibration video.

    • @MetrohmTV
      @MetrohmTV  ปีที่แล้ว

      You're welcome!

  • @bhagyaranasingha2216
    @bhagyaranasingha2216 7 หลายเดือนก่อน

    Nice explanation 👌

    • @MetrohmTV
      @MetrohmTV  7 หลายเดือนก่อน

      Thank you 🙂

  • @Radhe_Tyagi
    @Radhe_Tyagi 4 หลายเดือนก่อน

    Nice video...

    • @MetrohmTV
      @MetrohmTV  4 หลายเดือนก่อน

      Thanks!