hello i have a issue with a 2008t machine, i replaced acid pump because it wouldnt come up in conductivity it stays at 10.0, old pump i replaced brushes and it wouldnt come up either thats why i went ahead replaced with a new one but still not coming up. any suggestion
hello i have a issue with a 2008t machine, i replaced acid pump because it wouldnt come up in conductivity it stays at 10.0, old pump i replaced brushes and it wouldnt come up either thats why i went ahead replaced with a new one but still not coming up. any suggestion
@@kevinmunoz3389 Okay. Check valve 10 or 9 pressure transducer. If still Check valve 31. 33. 34. Or 32 one by one run dialysis mode