Agar aapne calibration unit laga rakhi hai to ref air dijiye otherwise koi jaroori nahi hai open karne ki . Aap sirf cal port me connect kijiye aur zero calibration kar sakte hain
Sir,your video is very helpful, pleas make a video for all parameter read and change and meaning full programming step by step in this zr 402g model, i hope early my request granted in your side sir.please.
Very well explained, but I am getting calibration error once I end the calibration after completing span and zero. Also, what are we supposed to do with reference port while giving Calgas?
Thank you for your video.! I have a question. After finishing calibration, the next step we must do for ensuring the system working accurately or not, Sir?
Thanks sir, Sample gas kon si hoti hai Aur reference gas kon si. Zero gas kon si hoti h Span gas kon si hoti h Aur cylinder ke connection kese hote hai sensor se?
Holding means output hold ho jata hai . Timing aap adjustment kar skate hain aapne according . Cross ka matlab samajh nahi aaya . Plz send me snapshot in my telegram group .
Sir , reference air Always continues flow in calibration time , in our plant 402g O2 installed in aux boiler then we are Always reference air continues in normal running plant ya normal service , we are not off reference air in calibration time.
Reference air should be continues flow . But not in my plant ,NO problem. Actually in my plant analyzers installed outside area so atmosphere air is enough and measur value is good . So didn't use in my plant .
Agar aapke paas 21% o2 standard gas cylinder available hai to best hai . Otherwise you know that atmosphere me bhi available hai 21% gas. So aap dono me koi bhi choose kar sakte hain
Sir ye Boiler me use hota , flue gas me oxygen concentration measure karne ke liye . Aur baanki Maine vedio me dikhaya hai . Koi aur querry hai to ask me .
Span Calibration 20.9 % me karte hain . Agar standard gas cylinder available nahi hai to aap atmosphere me bhi kar sakte hain . 8% standard se accuracy nahi aayegi .
I have open the cal port in zirconia Detector in atmospheric air & reference in port closed, but its oxygen reading is not increases in Zr402g display, what is the reason
Nice one 👌👌
Thanks 🙏🏻
Thanks for sharing your knowledge . Very helpful thanks
Thanks 🙏🏻
Well explained...
Thanks for your comment 🙏🏻🙏🏻🙏🏻
Well 👌
Thanks 🙏🏻
Thank you for your video 🙏🙏
Thanks 🙏🏻
Sir, mera important question hai ki O² Analyser me refrence gas kyu hota hai kab dena padta hai aur kitna dena padta hai
Well explain sir...
Thanks
Sir, jab zero calibration karte time, sensor ka reference port open rahega ya close ?
Agar aapne calibration unit laga rakhi hai to ref air dijiye otherwise koi jaroori nahi hai open karne ki . Aap sirf cal port me connect kijiye aur zero calibration kar sakte hain
Sir with sov and rotometer selection karte hai bateye
Sir,your video is very helpful, pleas make a video for all parameter read and change and meaning full programming step by step in this zr 402g model, i hope early my request granted in your side sir.please.
Ok sure I will try 🙏
Very nice
thanks for your comment
Very well explained, but I am getting calibration error once I end the calibration after completing span and zero.
Also, what are we supposed to do with reference port while giving Calgas?
And during the Hold Time, do we still have to supply the Calgas or we can turn off the knob?
Thank you for your video.! I have a question. After finishing calibration, the next step we must do for ensuring the system working accurately or not, Sir?
Yes obsoletely do . We can find out with the help of offline meter that our system is working properly or not .
Thanks sir,
Sample gas kon si hoti hai
Aur reference gas kon si.
Zero gas kon si hoti h
Span gas kon si hoti h
Aur cylinder ke connection kese hote hai sensor se?
Ok Mai aapke sabhi questions ko clear karne ke liye ek separate video banaunga .
Esko lagana kaise hai, O2 analyser ko, need instalation procedure
Thank you so much
Welcome 🙏🏻
Sir with rotometer and sov vedio banao
Please share calibration video for Emerson 6888xi o2 analyser
Calibration krty hove reference gas valve open ho ya colse?
Agar aap reference air use karte hain to open Rakhiye otherwise close rakhiye koi problem nahi hai .
Refrence gas rakhte kitna hai
Thanks bhai please exact location of application btaia means kon se line me kis unit me use ho rha h
Ok ✅ thanks for your suggestion
During calibration..... reference gas must be dummy or not???
Jab Zero ya Span calibration karte uss time refrence gas dummy karenge ya nahi
Sirr
Zirconia me pani Gaye toh kya hota sir kuch problem hota kya sir
Sir agar field display me hold and cross aa raha hai or 21 % dikha raha hai par dcs par zero to kya karu
Holding means output hold ho jata hai . Timing aap adjustment kar skate hain aapne according . Cross ka matlab samajh nahi aaya . Plz send me snapshot in my telegram group .
Good work, agr zero calibration me.0.4 ki bajaye 0.8 show ho Raha ho to os ko kaisy set Kia jaye through trimming??
Jab aap zero calibration karenge set value standard gas ke according to apne aap trimming le lega jab aapka calibration complete ho jayega
@@PowerPlantInstrumentation awo agr possible ho to is pa ak video bana ly.....
Thik hai 👍
Ges chemotherapy calibration shikhna hai
Sir cal time reference air should be open or not please reply
is not necessarily
@@PowerPlantInstrumentation thanks sir
Ye calibration plant jab running rahta hai tab karna hai....??? Ya fir band ke time karna hai???
@ jab bhi aapko need ho tab kar sakte hain .
Smoke detector ka video banao
If zero ratio too high how to resolution?
Nitrogen gas cylinder ke specifications kya hai
Sir , reference air Always continues flow in calibration time , in our plant 402g O2 installed in aux boiler then we are Always reference air continues in normal running plant ya normal service , we are not off reference air in calibration time.
Reference air should be continues flow . But not in my plant ,NO problem. Actually in my plant analyzers installed outside area so atmosphere air is enough and measur value is good . So didn't use in my plant .
Why used nitrogen gas for O2 calibration
Sir ETP plant he Dissolvid oxygen sensor in install karnahe
Aapko kya problem ho rahi hai install karne me
@@PowerPlantInstrumentation sir Dissolvid oxygen sensor se VFD speed very karnahe
Sir can we do the callibration in lab ( If we have both zero gas and callibration gas)
Of course
Span ka calibration krne k liye bhi cylinder use krna hota h ?
Agar aapke paas 21% o2 standard gas cylinder available hai to best hai . Otherwise you know that atmosphere me bhi available hai 21% gas. So aap dono me koi bhi choose kar sakte hain
@@PowerPlantInstrumentation bina zero calibration k bhi chala sakte h analyser ko ? Khali span calibration k saath
Haan chala sakte hain. Zero & span dono calibration karenge to accuracy good aayegi value ki .
Thanks
You are welcome 🙏🏻🙏🏻🙏🏻
Why are you doing span calibration first instead of zero calibration?
thanks
You are welcome 🙏🏻
Sir kitna me oxygen ko dena parta hai zero karne k lia
Rotamiter me kina pressure set karna pare ga
Please check Manual as per your analyser model
0.4%
@@skimran192270%
Cal zero/span air settings
Set calibration gas or reference air flow setting is 0.1 to 1.0 ltr/minute. If any other doubt then ask me on telegram @power plant instrumentation
Sir span port or zero port ka use nhi hai?
Cal and reference port hote hain . Cal port me zero and span gas connect karte hain calibration karne ke liye . Any other doubt ask me
How to troubleshoot, eror heating temp
check TC mv and Heater resistance
Calibration ke liye kahase 4%,,) cilindar kahade milega
Bharukha
Leas count anr tolerance kaise nikale
Sir hamara O2 analyzer showing 21% to 30% kaise isku solve kare
Zirconia cell voltage measure karo
Sir cell voltage -4.1mv
Thermo voltage 29.4mv
And Cell resistance is 177 homs
Can you translate in english
sir, I have problem. Display error 02 : heater temp. I can't use menu maintemance. please help sir
I don't understand your question.
Display error/heater temp/ ???
Please ask again
Calibration k time sensor ko boiler mein laga k hi rakhna h ya bahar nikalna padega ?
Aap kahin bhi kar sakte hain , Boiler me lagakar calibration karenge to best rahega sir .
Why did you calibrate zero gas at 0.50%?
Because 0% Oxygen concentrations cylinder not availed 👍
@@PowerPlantInstrumentationAlright
Koi yokogawa service mobile
Alarm11 TC voltage fault aa raha hai detector terminal 3,4 resistance check kiya to open as raha ,iska matalb
A11 cell voltage alarm hai so check cell and cell rubustness , clean cell terminals.
Replace cell which is inside the probe
Ye kaha use hota hai or kese kaam karta hai details me
Ans dijiye sir
Sir ye Boiler me use hota , flue gas me oxygen concentration measure karne ke liye . Aur baanki Maine vedio me dikhaya hai . Koi aur querry hai to ask me .
Sir Hamare yaha same aisa hi o2 system laga hua aur work o2 nhi bata rha hai
Display me koi error message ya alarm aa raha hai kya ?
Sir calibration pot open karne par o2 80% above dikha rha h kya reason hai.
Range check kijiye
Sir span cal 8.0 gas mai hota hai na?
Span Calibration 20.9 % me karte hain . Agar standard gas cylinder available nahi hai to aap atmosphere me bhi kar sakte hain . 8% standard se accuracy nahi aayegi .
Any other doubt / question so ask me
Sir O2 analyser 12% se increase and decrease nhi ho raha hai span and zero calibration karne par kyo
Standard gas sampling line check karo uske baad Zirconia cell replaced karke dekho
@@PowerPlantInstrumentation सर zirconia cell ठीक है या खराब कैसे चैंक करे
Reference in is open during calibration
is not necessarily
I have open the cal port in zirconia Detector in atmospheric air & reference in port closed, but its oxygen reading is not increases in Zr402g display, what is the reason
Check cell resistance and mv .
What is the use of reference port
Sir apka nam Kay hai
Hemant Singh