Floating Pool Monitor BLE-YC01 pH calibration at 7.0/4.0/10.0 pH calibrate 7.00 pH CAL
ฝัง
- เผยแพร่เมื่อ 15 ก.ย. 2024
- Open the YINMIK APP,
Setting- pH buffer- choose 7.00 4.00 10.01 (EU standard),
please ues 7.00 4.00 10.01 calibration solution
Then calibrate the device as shown in the video.
If choose 6.86 4.00 9.18 (Asia Standard)
Please use 6.86 4.00 9.18 pH solution
6.8/4.0/9.1 calibration procedure is the same as the 7.0/4.0/10.0 calibration procedure;
Lucia 0086 13166576796
Skype: toulucia2017
lucia@yinmikbest.com
It's not wise to move the monitor over the calibration solutions to wash the probe between measurements. Doing this carries a risk of dripping one solution into one of the others during the cleaning process. Better to place the cleaning solution in front of, or behind, the calibration solutions, so there is no risk of dripping from one beaker into the others.
BLE-YC01 calibration with MODE/CAL for 5+ secs does not work. How to reset PH to start a new calibration?
How calibrate EC and ORP?
The software is no so good. why can't I select the setpoint value to be calibrate ?
I bought this and by mistake run calibration when device was outside. Now I place it into a tap waters (~7) and run calibration but it think I'm calibration for 10 and not 7.
You have made a mistake, calibrated the pH in the air and caused the program to be disturbed. and it cannot be corrected.
pH calinration point is 4.00 6.86 9.18 or 4.00 7.00 10.01. it is standard, Calibration points cannot be set arbitrarily.
@@luciatou9203 too bad the software can't handle these error cases.
Hi, ich messe mit dem Smart Pool monitor 0.1mg/l die üblichen Teststreifen zeigen ein Ergebnis von 0.3mg/l.
Welches Ergebnis ist richtig?
MfG Michael
I calibrate the first 7.0 pH as shown on video, but when I put the sensor on 4.0 pH solution, it down only to 6.4 on display. And when I put it on 10 pH solution, it goes only until 7.6. Which means that I can't calibrate the 4.0 and 10.0 measurements. Is my sensor damaged? There is any way to reset the sensor?
This means that the electrode is broken. You need to contact your supplier to replace the electrode.
Thanks a lot
it's nothing
Hello, manual says use 250ml of water per beaker. In the video no more than 100ml is used. Which is correct?
It is to dissolve the powder with 250mL. But it does not mean that 250mL of solution must be used for calibration, the remaining uncontaminated calibration solution is stored for later use
How to reset all settings (calibrations)?
don't have this settings
How to calibrate CL?
After calibrating pH and ORP, CL is automatically calibrated
@@luciatou9203 Thank you. I already calibrated ph and orp. but sometimes CL doesnt show correct. just if has a lot cl on the water.
Hello
How to calibrate the EC and ORP?
EC use 1413μS/cm and 12.88 mS/cm calibration solution; ORP use is 222mV calibration solution
Hi Lucia, Does this device have the ability to test for Chlorine? And is there any way I Can use it to test for Alkilinty as well?
This model can test free chlorine. can't test the alkilinty
@@luciatou9203 cool! thanks for confirming!
Can the C-600 also test for Chlorine? Because I've read that the ORP is a measurement for checking Chlorine?