  • ELECTRICAL (At 23±5℃, 70% R.H. maximum)
    Efficient Measure Range:
    1mV/10mA: 10mA to 2 A, multimeter range is set to DC or rms AC 200mV
    1mV/100mA: 100mA to 20 A, multimeter range is set to DC or rms AC 200mV
    20A to 65 A, multimeter range is set to DC or rms AC 2V
    Precision:
    System Accuracy: The Precision of the current clamp add the Precision of the digital
    multimeter or scope.
    For example, if the conductor carries 100 mA current, the current clamp range will be set

ความคิดเห็น • 5

  • @rinopravato1285
    @rinopravato1285 7 หลายเดือนก่อน

    gracias

  • @7GIGEO7
    @7GIGEO7 ปีที่แล้ว +8

    Hey, even though i don't understand Spanish, i think you got a mistake in your calculations in 4:40. If your amplitude is 39mV then the Vpeak = 39/2 = 19.5V and your Vrms = Vpeak/Sqrt(2) = 19.5/1.414 = 13.8mV so the measured current with the Hantek clamp is I=Vrms * 10mA/mV =13.8mV * 10mA/mV = 137.8mA which is very close to the multimeter value which was 121mA and well within the spec of the current probe (+-1.5%+-5mV).
    I'm just saying because you might drive people away from a very reasonably priced and quite accurate current probe.

    • @javierhuarca9634
      @javierhuarca9634 ปีที่แล้ว

      Interesting. Perhaps the problem is the coupling? Because the oscilloscope Vrms is 42mV from first scope

  • @je1966
    @je1966 9 หลายเดือนก่อน +1

    3:49 No entiendo esa multiplicación de 41x10=480mA