Agilent GC Troubleshooting and Maintenance: Liner, Septum, and O-Ring Replacement

แชร์
ฝัง
  • เผยแพร่เมื่อ 23 ธ.ค. 2024

ความคิดเห็น • 20

  • @kavyasreechintalapudi
    @kavyasreechintalapudi 3 ปีที่แล้ว +5

    Could you please elaborate on the process of tuning the system after changing the liner and septum, including the parameters that should be checked to make sure everything is set?

  • @geoffstewart6617
    @geoffstewart6617 7 ปีที่แล้ว

    Way to go Herb!

  • @oskardumanski8538
    @oskardumanski8538 5 ปีที่แล้ว +2

    How should I clean the channel where the liner must be placed, if some particles are present?

    • @bygtyma05
      @bygtyma05 5 ปีที่แล้ว +4

      I use a long cotton swab and dip it in hexane or solvent. Focus on cleaning near the top of the inlet where the purge and split vent trap are. If you have particles on the gold seal, it's best to remove the column and either change out the gold seal or clean it with hexane. Best of luck...p.s. Herb and Daron taught me everything i know

    • @ПавлоКочубей
      @ПавлоКочубей 4 ปีที่แล้ว +1

      If you mean any particles are on the bottom of the inlet "tube" where the liner is placed, than just do the steps like in video of replacing the Gold Seal (without replacing, of course). It is pretty simple and requires two more steps than just replacing a column.

  • @eziobellavia2305
    @eziobellavia2305 27 วันที่ผ่านมา

    WHAT CAN I DO IF AFTER HAVING REPLACED THE LINER AND SEPTUM, THE TUNE TELLS ME THAT THERE IS A NITROGEN VALUE OF 38.3% AND OXYGEN OF 9.8%?

  • @aen8921
    @aen8921 8 ปีที่แล้ว +1

    How would you purge with carrier gas while the injector and column are still 'cold'?

    • @capchemist
      @capchemist 7 ปีที่แล้ว

      You can pressurize the inlet without needed to turn on any of the heating elements. This can be achieved from your instrument's front panel, or by adjusting the instrument parameters in your software.

  • @szamanisticrealistic2659
    @szamanisticrealistic2659 6 ปีที่แล้ว +1

    Is it necessary to vent the mass spectrometer to replace the inlet liner ?

    • @agilent
      @agilent  6 ปีที่แล้ว +4

      No, you don’t need to vent your mass spec to change your liner and septum. All you need to do is turn off the gas flow on your GC. As best practice make sure heated zones are cooled prior to the replacement.

  • @pharmasolutionpakistan7668
    @pharmasolutionpakistan7668 3 ปีที่แล้ว

    Good information

  • @nsy1087
    @nsy1087 5 ปีที่แล้ว

    For gcfid, do I replace the liner before or after the flame ignite

    • @angelicahall3701
      @angelicahall3701 4 ปีที่แล้ว

      Perform a liner replacement (and any other accessory replacements) prior to flame ignition.

  • @lycaetano
    @lycaetano 6 ปีที่แล้ว

    How can I replace the liner in GC with headspace?

    • @oskardumanski8538
      @oskardumanski8538 5 ปีที่แล้ว +1

      The same way as in the GC without HS:
      1.) Disconnect the transfer line (in modern systems transfer line is connected perpendicular to the liner and you don't need to disconnect it).
      2.) Remove the cap (in older systems it looks more like a sleeve than a cap).
      3.) Replace liner with accordance to the movie.

  • @jesusa1885
    @jesusa1885 6 ปีที่แล้ว

    How often should you do this?

    • @agilent
      @agilent  6 ปีที่แล้ว +4

      When to change the septum, liner ad O-ring depend on the analysis you are running how many injections are done a day, how clean the sample is, and what type of liner you are using. In January we did a Webinar on Healthy Habits of GC, through Separation Science. Here is a link. This should provide you with some good information.
      webinar.sepscience.com/habits-for-a-healthy-gc-insider-tips?hsCtaTracking=98fc9cbe-7a6d-429d-880f-01b5070d1e5f%7C2eb27d2d-ae36-4d74-b1f9-38302a368b7d

  • @gautammandrekar8
    @gautammandrekar8 2 ปีที่แล้ว

    Cotton wool baap dalega

  • @ХамедиллохонНурматхужаев

    Пожалуйста на русском языке