Troubleshooting and maintenance of ICP-MS Systems for optimal performance: Part 4: Interface Region

แชร์
ฝัง
  • เผยแพร่เมื่อ 8 พ.ค. 2017
  • Dirty, damaged, or blocked interface cones drastically impact the sensitivity and background of your ICP-MS system. Small/gradual changes in the vacuum level of the interface region is an indicator of cone maintenance. Learn about how to clean and condition your interface cones, as well as more information about maintaining extraction lenses.
    Proper cleaning and maintenance techniques can extend the lifetime of your Agilent ICP-MS system and avoid instrument drift. In this video, we provide steps and tips to help you with cone cleaning and troubleshooting. This video is part 4 of a four-part series covering routine preventative maintenance of ICP-MS Systems to improve product lifetime and result confidence.
    👉 For information on Agilent ICP-MS Sampler Cones, please visit: www.agilent.com/en/product/at...
    👉 For information on Agilent ICP-MS Skimmer Cones, Skimmer bases, and Interface Cone Supplies, please visit: www.agilent.com/en/product/at...
    👉 For more information on ICP-MS Maintenance & Troubleshooting, please visit: explore.agilent.com/icp-ms-re...
    👉 For ICP-MS Systems & Supplies, please visit: www.agilent.com/en/product/at...
    👉 To watch Part 1 on standard and sample preparation, diagnosis, and maintenance guidelines, please visit: • Troubleshooting and ma...
    #ICPMSMaintenance #ICPMSCones #Smaplercone
  • วิทยาศาสตร์และเทคโนโลยี

ความคิดเห็น • 6

  • @A_Dopamine_Molecule
    @A_Dopamine_Molecule 6 ปีที่แล้ว +3

    I am training for a job that requires mastery of Agilent MassHunter quant and qual software, and I am working with reverse-phase HPLC/tandem MS machines. Do you have any videos that can supplement my training? It would be a great help, thanks.

  • @romasinghal123
    @romasinghal123 2 ปีที่แล้ว

    How to know if there is sample aspiration problem??

  • @mohammed19812009
    @mohammed19812009 7 ปีที่แล้ว

    Can you show how to optimize the system by tunning

    • @agilent
      @agilent  7 ปีที่แล้ว

      Tuning the Agilent ICP-MS instruments is made simple and automated with Auto Tuning. This ensures the instrument is always optimized, independent of the operator.
      There is a video available on our eFamiliarization guide provided with all Agilent ICP-MS. You will find the “Tune Overview” video under MassHunter Basics in the menu.

    • @agilent
      @agilent  6 ปีที่แล้ว

      For organic analysis it is recommended to ignite the plasma in organic mode while aspirating isopropanol (IPA). Tuning and optimization of plasma parameters, lens parameters and ORS parameters should be performed by aspirating a 1 ppb tuning solution prepared from an inorganic Stock Tuning solution (Part No. 5188-6564) diluted in IPA. Please see relevant application note for further information: www.agilent.com/cs/library/applications/5991-3538EN.pdf

    • @MaxPowersLTD
      @MaxPowersLTD 6 ปีที่แล้ว

      Also keep an eye on the reflective power before performing the hardware tune. Perform an auto rf matching if it is high. The shunt capacitor may need to be recalibrated for lower reflective power.