Common Mistakes in Autoclave Validation ​⁠

แชร์
ฝัง
  • เผยแพร่เมื่อ 9 ก.พ. 2025
  • Join this channel to get access to perks:
    / @pharmaven
    Why Air Removal is Required in Autoclave?
    How to calculate F0 Value?
    What is Overkill Cycle in Steam Sterilization?
    #autoclavevalidation
    #Validation
    #Validationinpharma
    #sterilization
    #autoclave
    #aseptic
    #usfda
    #steam
    #bacteria
    #bacterial
    #filter
    #pharmaknowlege
    #sterility
    #qualityassurance
    #quality
    #sterile
    #gmp
    @PHARMAVEN ​

ความคิดเห็น • 3

  • @patelvishal2313
    @patelvishal2313 6 วันที่ผ่านมา +1

    Thank you sir..

  • @NiteshKohad
    @NiteshKohad 5 วันที่ผ่านมา +2

    Sir how to determine the slow to heat and during which stage it should be determine

    • @PHARMAVEN
      @PHARMAVEN  5 วันที่ผ่านมา +1

      Slow to heat location can be determined by placing multiple sensors in empty chamber.
      Slow to heat load items can be determined by placing sensors in all different types of load items and evaluating temperature data to understand which load item took most time to heat up to sterilization temperature