The video does not talk about needle alignment. Make sure, to align needle and adjust height of the needle using Lab Advisor for a successful change of needle. The manual talks about the procedure.
Dear Customer, Thank you for your request. The needle clamp screw (0515-2606) and the arm spare kit (G1313-68713), which contains not only two screws but also the clamp plate (G1313-24101), can be ordered in our online store [www.chem.agilent.com/store/]. You are correct, the information is not provided in the User Manual, but can be found in the corresponding Service Manual. We apologize for the inconvenience. Please don’t hesitate to contact our support [www.agilent.com/en/contact-us/page] in case of any further questions.
The video does not talk about needle alignment. Make sure, to align needle and adjust height of the needle using Lab Advisor for a successful change of needle. The manual talks about the procedure.
How to get the needle screw, no part number for that item on the user manual of the autosampler.
Dear Customer,
Thank you for your request. The needle clamp screw (0515-2606) and the arm spare kit (G1313-68713), which contains not only two screws but also the clamp plate (G1313-24101), can be ordered in our online store [www.chem.agilent.com/store/].
You are correct, the information is not provided in the User Manual, but can be found in the corresponding Service Manual. We apologize for the inconvenience. Please don’t hesitate to contact our support [www.agilent.com/en/contact-us/page] in case of any further questions.
after replacing needle need to do alignment without alignment needle will get bend