  • @marcoaguiari3591
    @marcoaguiari3591 ปีที่แล้ว +1

    I test Vacuum, Helix e Bowie Dick vanno eseguiti tutti i giorni in cui si usa l'autoclave ( EN 13060 : 2019), QUESTO FILMATO VA AGGIORNATO.

    • @TecnoGazIndustries
      @TecnoGazIndustries  ปีที่แล้ว

      Buongiorno, le indicazioni sono generali in quanto le normative cambiano a seconda delle regioni. Cordiali saluti

  • @RosaParis-y4v
    @RosaParis-y4v ปีที่แล้ว

    Come vengono smalite le provette utilizzate per il test biologico?

  • @karmenfichera75
    @karmenfichera75 ปีที่แล้ว

    Dove si acquistano i manuali per l'archiviazione?

    • @TecnoGazIndustries
      @TecnoGazIndustries  ปีที่แล้ว

      Buongiorno,I manuali per archiviazione (Sterildocs) si acquistano presso i dealer che vendono le apparecchiature (locali o nazionali)

  • @annina8114
    @annina8114 3 ปีที่แล้ว +1

    Ma quindi helix o b&d si equivalgono, va fatto l'uno o l'altro non entrambi..

    • @Bogdan_Viktor
      @Bogdan_Viktor 3 ปีที่แล้ว +2

      Non è uguale, helix test è per i corpi cavi, come manipoli, mentre il b&d test è per i corpi porrosi, come per esempio camici

    • @TecnoGazIndustries
      @TecnoGazIndustries  ปีที่แล้ว

      Buongiorno, Helix e B&D verificano 2 attitudini diverse dell’autoclave:
      Helix verifica la capacità di penetrazione del vapore nei corpi cavi di tipo B; B&D verifica la capacità di penetrazione del vapore nei corpi porosi.
      Vanno fatti entrambi, in momenti diversi (non mettere i 2 test insieme nello spesso ciclo …anche se il testo è identico e corrisponde ad un solo tasto sulla plancia autoclave)