GC - Gas Chromatography - Split & splitless injection Animation HD

แชร์
ฝัง
  • เผยแพร่เมื่อ 16 พ.ย. 2018
  • Support and hit like and/or subscribe =). Thank you!
    A basic video about the concept of the split and splitless injection. Good video for beginners.
    Learn the basic concept of Gas Chromatography
    1. introduction
    • GC - Gas Chromatograph...
    2. Mobile phase of GC
    • GC - Gas Chromatograph...
    3. injection and vaporization of sample (this video)
    4. GC Capillary column
    • GC - Gas Chromatograph...
    5. GC - FID - Flame Ionization
    • GC - Gas Chromatograph...
    and more to come.

ความคิดเห็น • 29

  • @thukiRC97
    @thukiRC97 5 ปีที่แล้ว +2

    Really helpful, thank you!

  • @pharmaboy1510
    @pharmaboy1510 3 ปีที่แล้ว +2

    It is actually simple science. Thank you

  • @RajnishKumar-hx4qg
    @RajnishKumar-hx4qg 2 ปีที่แล้ว +2

    This is what Mr. Simple Science...Clearly and easily understood...👍

  • @hamidali5236
    @hamidali5236 5 ปีที่แล้ว

    Tremendous one!!

  • @uchihaangel1619
    @uchihaangel1619 3 ปีที่แล้ว

    Thank you very much. It 's a good video.

  • @007imranjan
    @007imranjan 4 ปีที่แล้ว

    Well done..👍

  • @ujjwalthorat2020
    @ujjwalthorat2020 5 ปีที่แล้ว

    Really very helpfull..

  • @EyeAngelo
    @EyeAngelo 4 ปีที่แล้ว +1

    Clear explanation!!! Thank a lot

  • @riyanshikacollections8925
    @riyanshikacollections8925 3 ปีที่แล้ว

    gud explanation..thank u

  • @Authorofpity
    @Authorofpity 4 หลายเดือนก่อน

    This is bestest!!!

  • @biometal770
    @biometal770 5 ปีที่แล้ว +2

    Yes good video. The liner to column interface was one of constant wonder for me, whether it seals against the column and such. Some liners make a decent seal, some do not, depends on the liner.

    • @bygtyma05
      @bygtyma05 4 ปีที่แล้ว

      The liner should sit flush against the gold seal.There is a small channel on the surface of the gold seal that allows your carrier gas and the sample to pass through.

  • @vaibhavsingh1193
    @vaibhavsingh1193 4 ปีที่แล้ว +1

    it worked i my exams!!

  • @matteodiogene3084
    @matteodiogene3084 3 ปีที่แล้ว

    cool

  • @uthpalaruvini7324
    @uthpalaruvini7324 3 ปีที่แล้ว +1

    Content is really helpful.one more suggestion.could you please add sound to video.I mean read them

  • @mawazeeno
    @mawazeeno 5 ปีที่แล้ว +1

    The video is very helpful,thanks for that. But what do you mean by , make sure the sample does not exceed the liner, please explain this part . thank you ?

    • @biometal770
      @biometal770 5 ปีที่แล้ว +3

      The volatilized sample volume should not exceed the volume of the liner. You might be able to get away with it, as there is additional volume in the inlet. Also, the gas is constantly flowing, if you did inject a large enough volume to force some sample into the gas line, it would just flow right out through the column. You would see it in your chromatogram as a severly overloaded column lol.

  • @syedmarjuk662
    @syedmarjuk662 2 ปีที่แล้ว

    Great video to understand. But im having very less quantity of sample, which spilt ratio I have to go?

  • @yulianianisamayapradipta5474
    @yulianianisamayapradipta5474 2 ปีที่แล้ว

    what is the meaning if i see in the journal he/she uses split-split less injector. is he/she uses split injector/split less injector/ or both?

  • @shoaibrazakhan4198
    @shoaibrazakhan4198 5 ปีที่แล้ว

    that really did help . but, according to how u you defined split ratio that ratio of amount of gas flowing to column to that of flowing to split valve, so in that case higher ratio means higher amount of sample to the column that will surely overload the column and consequentially less accurate results will be produced. ratio must be like, ratio of amount of sample sent to split line to that of sent to column. so in this case higher ratio means small amount of sample sent to column. is this so???

    • @bygtyma05
      @bygtyma05 4 ปีที่แล้ว

      higher ratio will mean more of the sample will go out to the split than into the column. This will result in smaller peaks

  • @yu9234
    @yu9234 ปีที่แล้ว

    2:29 shouldn't it be splitflow/column flow?

  • @ibrahimbelal6904
    @ibrahimbelal6904 3 ปีที่แล้ว

    Thank you for smart video, I prepare material about Gas Chromatography can I use this video in my material?
    please if you approve I'll be appreciate for you also I'll mention you page in my material.

  • @faithwangui7200
    @faithwangui7200 2 หลายเดือนก่อน +1

    I thougt my computer sound settings were off lol

  • @apnacoffeebreak7427
    @apnacoffeebreak7427 3 ปีที่แล้ว

    Why do we use mostly Split injection in GC?

    • @thegingerbreadman
      @thegingerbreadman 3 ปีที่แล้ว

      1. To prevent overloading the column. 2. In some cases, it produces sharper more well define peaks

  • @samikshasalunke4656
    @samikshasalunke4656 4 ปีที่แล้ว

    Awaj ka nahi yet

  • @user-pk2wb6js4u
    @user-pk2wb6js4u 8 หลายเดือนก่อน

    sorry...there is no Voiceover