    Waht is the different between Ad11 and ad 12?

    • @ADWAHungary
      @ADWAHungary  11 หลายเดือนก่อน

      Dear Kamran Pipekesh,
      The difference between the testers is that AD11 is one decimal, AD12 two decimal accurate tester.
      Best regards,
      Adwa Team

  • @facparis7064
    @facparis7064 4 ปีที่แล้ว +1

    Hi. I tried to do it like that but after 7.01 appears 10.01. Do you know why?

    • @ADWAHungary
      @ADWAHungary  2 ปีที่แล้ว +1

      Dear Fac,
      It could be a lot of things, E.g. the calibration solution was old or contaminated; the bulb is dried out; etc.
      Please send your question/request to support@adwainstruments.com to help you.
      Thank you!

    • @nadreenshandgawe8383
      @nadreenshandgawe8383 ปีที่แล้ว

      @Adwa Instruments I wont to buy
      How?

    • @arturoairsoft235
      @arturoairsoft235 10 หลายเดือนก่อน

      ​@@ADWAHungaryviene con el líquido del bulbo?y cuantos líquidos de calibración?es mi primera vez comprando uno y aunque no tenga mucho dinero tal vez sea buena idea pedir los líquidos a parte también

    • @ADWAHungary
      @ADWAHungary  10 หลายเดือนก่อน

      @@arturoairsoft235 The meter is supplied with a protective plastic cap (with stroage solution in it) and one pH 7.01 and 4.01 20 ml sachet for calibration.
      You can see all the details and additional accessories and solutions on this link: adwainstruments.com/pocket-testers/103-waterproof-pocket-testers/professional-waterproof-ph-temp-pocket-testers/83-ad12

    • @diego.zaisov
      @diego.zaisov 10 หลายเดือนก่อน

      Its, becouse there is no way to calibrate if is too far from the buffer point. Its done, buy another one

  • @gj1706
    @gj1706 ปีที่แล้ว

    ประเทศไทย​ มีตัวแทนจำหน่าย​ไหมครับ

    • @ADWAHungary
      @ADWAHungary  ปีที่แล้ว

      Dear @gj1706,
      You can find contact details to our distributor in Thailand on our website: adwainstruments.com/index.php?option=com_idj
      Best regards,
      Adwa Team

  • @hanneshenning
    @hanneshenning ปีที่แล้ว

    When I calibrate and he wants me to use 7ph. I put it in and immediately it's jumps to 4ph rec... Can't make it work it jumps back and forth from 7 to 4 and only is possible to calibrate a 7ph on 4ph... New product first calibration... sad af

    • @ADWAHungary
      @ADWAHungary  ปีที่แล้ว

      Dear Hannes Henning,
      We are sorry for hear that.
      For your problem, we can offer the following solutions: probably the probe has dried out.
      Firstly, try to measure the pH 4.01 buffer solution without calibration.
      If the value is more or less +/- 0,5 pH than the expected (4.01pH ) value, you need to reactivate the probe.
      Soak the tip of the probe in buffer solution (or pH4 solution) for a few hours to reactivate/hydrate it.
      After try to calibrate again.
      We would like to help you to have a good, useful product. For further assistance, please contact us via email.

    • @hanneshenning
      @hanneshenning ปีที่แล้ว

      @@ADWAHungary I did and it do not work... It always jumps from 'use 7' to 'rec 4' the probe has never been dried out I kept it wet always. Its not fun

  • @diego.zaisov
    @diego.zaisov 10 หลายเดือนก่อน

    HOW TO SWITCH FROM 1 Point Calibration to 2 Point Calibration!!! please this information is nowhere to be found

    • @ADWAHungary
      @ADWAHungary  10 หลายเดือนก่อน

      Dear Diego,
      AD11, AD12 and AD14 are automatically set up for two-point calibration.
      For 1-point calibration:
      For a one-point calibration with buffers pH 4.01 or 10.01, the meter automatically accepts the calibration when the reading is stable. Then it will show „OK 1”
      If a one-point calibration with buffer pH 7.01 is desired, then after the calibration point has been accepted, press the ON/OFF/MODE button to return to normal mode. Then it will show „OK 1”
      For 2-point calibration:
      For a two-point calibration, place the electrode in pH 7.01 buffer. After the first calibration point has been accepted, the “pH 4.01 USE” message appears. If it accepts the pH 4.01, then it will show „OK 2” message.
      You can see the full description in the manual: adwainstruments.com/manuals/manual_ad11_12.pdf
      If you have more question, please send us an email: adwainstruments.com/contact-us/contact-adwa-office

    • @diego.zaisov
      @diego.zaisov 10 หลายเดือนก่อน

      ​@@ADWAHungary No its not. When i put the ph meter in to the 7.01 solution its shows 10.1 REC and thats it.

    • @ADWAHungary
      @ADWAHungary  10 หลายเดือนก่อน

      @@diego.zaisov Please, check the value of 7.01 pH buffer solution.
      Probably the measuring value is too far from the expected 7.01 pH, and closer to 10.01pH and the meter automatically ask for 10.01.
      In this case try the following:
      - reset the tester -see the instruction in the manual
      - Soak the tip of the probe in storage solution for at least few hour to rehydrate it.
      - replace the probe