What is the disinfectant load and What is the warm up cycle What is the hot leck test What is the sip filter What is the filter integrity What material required for autoclave HPHV prosess Steam sterilizer
Disinfectant load - item used for Disinfectantion and cleaning Warm up cycle - simply to heat the Autoclave chamber Hot Leak Test - leak test done after warmup cycle or empty cycle, leak test in hot chamber SIP of Vent Filter by steam Sterilization Filter Integrity - to check that filter is free from damages and blockages, it should be integral HPHV process - involves high pressure and high vacuum pulses to achieve Sterilization You can watch our more videos on steam Sterilization process to understand better
No, it should be qualified as per the pre determined process parameters or else you need to develop the qualification cycles prior to qualify any load.
Yes you can do it but first you need to evaluate the impact of high temperature on the materials which you are sterilize. An increase in temperature will reduce the time duration.
Aap bahut achhe se samjhate hai sir
Thanks dear
Thank you ,good demonstrated.
Thank you
What is the disinfectant load and
What is the warm up cycle
What is the hot leck test
What is the sip filter
What is the filter integrity
What material required for autoclave
HPHV prosess
Steam sterilizer
Disinfectant load - item used for Disinfectantion and cleaning
Warm up cycle - simply to heat the Autoclave chamber
Hot Leak Test - leak test done after warmup cycle or empty cycle, leak test in hot chamber
SIP of Vent Filter by steam Sterilization
Filter Integrity - to check that filter is free from damages and blockages, it should be integral
HPHV process - involves high pressure and high vacuum pulses to achieve Sterilization
You can watch our more videos on steam Sterilization process to understand better
Do we need to qualify sterilizer as per production temperature or 1°C less than production temperature?
No, it should be qualified as per the pre determined process parameters or else you need to develop the qualification cycles prior to qualify any load.
@@PharmaGyan During the cycle development which one we need to consider from the above two?
Production temperature
Auto clave me vlt ka pass minute kya hai
👍👍👍👍
Auto clave me vacuum fail ho jaane par kya karte hai
Check the ports sealing gaskets etc which may lead to air infiltration
Brother how to stop in middle of this machine
Pvl
Auto clave me vlt two minute se zyada aa raha hai to kya karana padta hai
2 minute se jyada matlab
VLT measurements time me ni hota pressure me hota h
Clarify your question dear
My question is why only general standard temp for steam sterilization is 121•c ,
But if we use more than that it will do Complete our process fast
Yes you can do it but first you need to evaluate the impact of high temperature on the materials which you are sterilize.
An increase in temperature will reduce the time duration.
Thank you
th-cam.com/users/shortsCibCM-t9zDY?feature=share
Why 121.1 required..
Its a std temperature you can change as per your requirements
हिंदी में नही।हैं
D-Value samaj nhi aaya
We will try to prepare a separate video on D value along with calculation of D value
Bottles