No.103 - Fluke 43B Power Quality Analyzer Repair

แชร์
ฝัง
  • เผยแพร่เมื่อ 26 ส.ค. 2024
  • ** PCBWAY For all your PCB Services at www.pcbway.com... **
    I thought this 43B would be useful around the workshop and bought it faulty. Turns out it make be beyond fixing!
    #electronicscreators
    Visit my website for lots more - www.ianjohnsto...
    Help me make more of these videos:
    Pledge using Patreon - / iansjohnston
    Donate using PayPal - www.paypal.me/...
    Follow me on Twitter - / iansjohnston

ความคิดเห็น • 44

  • @mr1enrollment
    @mr1enrollment ปีที่แล้ว +6

    actually I am amazed that you fix many of the instruments that you find.
    given the level of integration in modern stuff - any win is a big win.

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +5

      The YT video foreshortens the work that goes on in the background. Some of the repairs take days/weeks and a good bit of thinking and pouring over the schematics and hardware. But usually, there's a point where I go "ahhhhh, thats gotta be it"......then I hit record again.....😀

  • @xDevscom_EE
    @xDevscom_EE ปีที่แล้ว +8

    Back when manufacturers were nice and provided service information and schematics for products they made for people....

  • @davidv1289
    @davidv1289 ปีที่แล้ว +5

    Let's hope you can fix it by re-flowing the flash ROM. The 43 is an excellent tool, I used one routinely to check 75 KW 480 volt 3 phase UPS/PCS systems for harmonic distortion. Regards, David

  • @feedback-loop
    @feedback-loop ปีที่แล้ว +6

    There is a chance that the problem might be with the keypad. I have seen Fluke 19x scopemeters, which are very similar, not powering up because of some shorts in that flexible keypad membrane. Try disconnecting the keypad altogether and powering up the unit by shorting the power button line (marked ON KEY) to the ground.

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +1

      I checked that, C400 on the Pcb just needs shorted out as it is across ON KEY and 0vdc.

    • @feedback-loop
      @feedback-loop ปีที่แล้ว +1

      @@IanScottJohnston Did you disconnect the keypad?

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +1

      Yes.

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +19

      STOP PRESS.....I have just reflowed the Flash......lets say there should be another video very soon....:)

    • @TheEmbeddedHobbyist
      @TheEmbeddedHobbyist ปีที่แล้ว

      @@IanScottJohnston How the hell did you find something to flash into it?

  • @BradleyMabe
    @BradleyMabe ปีที่แล้ว +1

    Very cool step-by-step logic troubleshooting - I enjoyed the way you took us through it - It would be interesting to see if this is a common failure mode for this Fluke product ... Thanks!

  • @allthegearnoidea6752
    @allthegearnoidea6752 ปีที่แล้ว

    I bough a fluke power analyser data logger last month. Downloads into a laptop. Very useful for tracking down issue with the customers power supply.

  • @elylopes7343
    @elylopes7343 ปีที่แล้ว

    I have one Fluke 434 Power Quality Analyzer showing the same issue that you described in this video. I sent it for repairing to the Fluke representative here in Sao Paulo, Brazil a few months ago, but they returned it to me saying they could not fix it up because it was obsolete and Fluke did not hold parts for it any more.

  • @BjornV78
    @BjornV78 ปีที่แล้ว +2

    You can try to heat up the flashrom and Asic chips by hot air , and see if that gives a difference.

  • @t1d100
    @t1d100 ปีที่แล้ว

    Nice review of the fault.

  • @andymouse
    @andymouse ปีที่แล้ว

    Great trouble shooting and watch this space !....cheers.

  • @SidneyCritic
    @SidneyCritic ปีที่แล้ว

    I've seen quite a few repairs were those flat ICs crack solder joints.

  • @ickipoo
    @ickipoo ปีที่แล้ว +1

    I would think that, for the ASIC to be outputing the watchdog signal, a fair bit of code has to be loaded and running?

  • @MR_Garage
    @MR_Garage 5 หลายเดือนก่อน

    I read below that you re-flowed the flash and all is working again? Just the screen correct having lines (common issue)? Replacement screens on ebay I have found.

  • @andersvandegevel8355
    @andersvandegevel8355 ปีที่แล้ว

    That would be a flow-chart rather than a truth table, but in any case, hats off, if you can't fix it, I doubt anyone can

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +1

      Yep....dunno why I said truth table....LOL!

    • @andersvandegevel8355
      @andersvandegevel8355 ปีที่แล้ว

      @@IanScottJohnston a day without at least one brain-fart is misspent!

  • @elsaarmstrong-zp6ng
    @elsaarmstrong-zp6ng ปีที่แล้ว

    I thought all modern power adaptors had the centre contact positive ?

  • @proluxelectronics7419
    @proluxelectronics7419 ปีที่แล้ว

    So much gear has Flash based firmware, It's a good way for manufactures to put a time limit on the life of a unit, the flash just hangs on a bad block. I have a few nand failure project/repairs with very little information on nand rebuilding, unless you know Linux.

    • @TheEmbeddedHobbyist
      @TheEmbeddedHobbyist ปีที่แล้ว

      We tried to claim that the code in an ASIC flash was firmware, but the power's that be were trying to insist that it was software and should come under all the rules and standards for software production. it did go in our favor for that project but they got their way in the end.

  • @amirb715
    @amirb715 11 หลายเดือนก่อน

    Did you have another go at this Fluke 43B? If mask software is running it should be possible to flash the FW through the optical-serial port.

    • @IanScottJohnston
      @IanScottJohnston  11 หลายเดือนก่อน

      Not yet, I have it on the back burner to get back to. Thanks for the info!

    • @amirb715
      @amirb715 11 หลายเดือนก่อน

      @@IanScottJohnston read sec 7.4.1,7.4.2 of service manual. Although there is a dead end when it says look at sec 7.6 and that section does not exist in the published manual 😞

  • @elylopes7343
    @elylopes7343 ปีที่แล้ว

    Have you fixied the issue witl the flash memory? Did you try resoldering it to the PCB?

  • @user-xr5el7tk6i
    @user-xr5el7tk6i 9 หลายเดือนก่อน

    There are schematics fluke 434 ?

  • @user-xr5el7tk6i
    @user-xr5el7tk6i 9 หลายเดือนก่อน

    There is a schematic fluke 434?, won't turn on.

  • @fredflintstone1
    @fredflintstone1 ปีที่แล้ว

    you can't win them all but maybe just maybe there will be a part 2🙂

  • @piconano
    @piconano ปีที่แล้ว

    Can you not de-solder it and send it to Fluke and get a new replacement?
    It's very rare for ICs to go bad all of a sudden after operating for months.
    Specially on a battery operated instruments that eliminate the possibility of AC surges.

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +4

      It's working now. I reflowed the flash and it's now powering up. The LCD on the other hand has issues, so will be repairing that.

    • @piconano
      @piconano ปีที่แล้ว

      @@IanScottJohnston Excellent. Hope to see it working soon.

  • @timturner7609
    @timturner7609 ปีที่แล้ว

    11:56 Nobody puts relay in the corner!

  • @TKomoski
    @TKomoski ปีที่แล้ว

    Is it really worth chasing a fix ?

    • @IanScottJohnston
      @IanScottJohnston  ปีที่แล้ว +6

      For YT, the fun & satisfaction of it and not to mention my OCD of it sitting on a shelf somewhere un-fixed would bother me LOL!

    • @TKomoski
      @TKomoski ปีที่แล้ว

      @@IanScottJohnston Yes I have seen you replace almost every componet on some of your repairs. I guess if you realy need that functioning then a repair is in order. Don't get me wrong Ian I admire your dedication to effect a repair. Cheers

  • @radio-ged4626
    @radio-ged4626 ปีที่แล้ว

    Well, you're animated figure isn't waving a white flag at the end of the video so maybe there's hope 🙂