I found this note about the high reflection error in the User Manual, hope it helps! This message occurs when the connection to the test port produces large reflections back into the instrument. This can compromise measurement accuracy. To fix this error, check all junction connections, or remove the RF connection or component that is causing the high reflection. If a cable or antenna is connected, then it could be an indication of poor RF transmission capability.
I found this note about the high reflection error in the User Manual, hope it helps!
This message occurs when the connection to the test port produces large reflections back into the instrument. This can compromise measurement accuracy. To fix this error, check all junction connections, or remove the RF connection or component that is causing the high reflection. If a cable or antenna is connected, then it could be an indication of poor RF transmission capability.
The issues is not at 8.45 ft.... you must take the peak, in that case you got two faults : first connector and the last connector. try to torque...
Yep 115.91 dB at 18.82 ft is the peak.
I am getting a high reflection error when I am starting calibration. Any thoughts?
Either you pim load or pim standard is faulty