Instrumentation for Nanosecond and Femtosecond Transient Absorption Spectroscopy - Part 2

แชร์
ฝัง
  • เผยแพร่เมื่อ 4 ม.ค. 2025

ความคิดเห็น • 6

  • @artisharma4313
    @artisharma4313 ปีที่แล้ว +2

    Beautifully explained the whole instrumentation of TA spectroscopy

  • @adamroach3771
    @adamroach3771 ปีที่แล้ว +3

    Awesome lecture!

  • @DSeyit
    @DSeyit ปีที่แล้ว

    Hi, do you know how the phase of the differential signal is chosen? Because sometimes ccd will measure on/off/on/off data and sometimes off/on/off/on data which is flipped. I am having hard time to fix this issues.

    • @tintobress
      @tintobress ปีที่แล้ว

      Hi, you need to measure the phase of your chopper together with the first spectrum acquired by your CCD. once you know if the first shot corresponds to a "pump open" or a "pump closed" event you are all sorted. Here I am assuming that chopper and CCD are synchronised to the laser

  • @yizhu5275
    @yizhu5275 2 ปีที่แล้ว +1

    apparently the time of light traveling 1cm is calculated incorrect. 1cm should be 0.033 ns rather than 0.33 ns.