Good review of the standard. But there are some questions might be raised around usage of PCDs for PQ. Manufacturers use real inoculation methods under overkill approach. However trying to do similar practice in hospitals is time consuming (1 week for incubation period) and extremely costly. After each major service or maintenance performing PQ is not sustainable. PCDs can be used for PQ which is not mentioned in this video (as far as I follow without Spanish). For external labs it will be a great challenge for them to perform these tests for VH2O2 sterilizers as they mainly practiced according to ISO 14937 in the past.
La exposición en relación a la Norma ISO 22441, es muy clara tanto por la terminología utilizada como por la organización de la exposición. Muchas gracias colega
Good review of the standard. But there are some questions might be raised around usage of PCDs for PQ. Manufacturers use real inoculation methods under overkill approach. However trying to do similar practice in hospitals is time consuming (1 week for incubation period) and extremely costly. After each major service or maintenance performing PQ is not sustainable.
PCDs can be used for PQ which is not mentioned in this video (as far as I follow without Spanish).
For external labs it will be a great challenge for them to perform these tests for VH2O2 sterilizers as they mainly practiced according to ISO 14937 in the past.