The radiography of the foot demonstrated 3.42 is not correct. First, the global standard for and AP is with a 15 degree uptilt, and second, we see that no collimation was used, so the lower leg was included in the beam which added to dose and scatter degradation of the radiograph. There should be no reason for a radiographer to wear a lead apron when doing a chest xray or a foot xray.
Bibinagar AIIMS pe kya IVF treatment hota hain? Please bataiye
Woww🥳
The radiography of the foot demonstrated 3.42 is not correct. First, the global standard for and AP is with a 15 degree uptilt, and second, we see that no collimation was used, so the lower leg was included in the beam which added to dose and scatter degradation of the radiograph. There should be no reason for a radiographer to wear a lead apron when doing a chest xray or a foot xray.