  • @Runco990
    @Runco990 7 หลายเดือนก่อน +2

    I don't leave my 53131A powered on all the time. While it IS calibrated, I run it on my GPS/Rubidium standard. I very much remember the 90's capacitor plague. Edit: I DID change the fan to a noctua mounted with rubber mounts. It is whisper quiet. I really hate the whiny fans that have shown up in everything so they can make the tiniest heat sinks. Worthwhile mod for anyone that leaves these counters always connected.

  • @PapasDino
    @PapasDino 7 หลายเดือนก่อน

    I leave my 131A unpowered since I don't routinely use it and since I have a Rb standard I turn both on when I need it since for my ham workbench that's fine. I do have a knock-off after market mid-level oven but haven't mounted it yet. You convinced me I need to investigate my counter for recapping, thanks! 73 - Dino KLØS

  • @NearFarMedia
    @NearFarMedia 7 หลายเดือนก่อน

    I have had to recap both my 53132A and a 53181A (which went to my brother's lab). The caps in the PSU's were well dead and leaking. In the 53181A even some of the ceramic caps were bad too surprisingly...
    Both are now running great. I updated the firmware and replaced the fans and used rubber mounts to get them all nice and quiet again, as well as replacing the dimming VFD displays with new bright ones sourced from ebay.
    How are you performing the fine time interval calibration? Or didn't you bother with that?

    • @ZenwizardStudios
      @ZenwizardStudios  7 หลายเดือนก่อน

      I just did the time base alignment. There are 3 processes Quick , Time base, and Fine at the moment I do not have the equipment to generate the fine required waveform. I left the calibration offsets alone and just the time base alignment.

  • @SQ5DBF
    @SQ5DBF 7 หลายเดือนก่อน

    Nice video. Ripple and noise(bad caps) on power supply rails +5/+12/-12V can affect timebase stability.

  • @jspencerg
    @jspencerg 7 หลายเดือนก่อน

    Makes me want to inspect/ recap all the 90's units...

  • @sebastian_harnisch
    @sebastian_harnisch 7 หลายเดือนก่อน

    Do you happen to have a list of the capacitors required for a full recap? I have an identical unit, also with the HS oven, would be really handy ;)

    • @Vitalii-R.
      @Vitalii-R. 7 หลายเดือนก่อน +2

      I have, try to find little bit later

    • @ZenwizardStudios
      @ZenwizardStudios  7 หลายเดือนก่อน

      I don't have a list per say. Ill be honest I have a giant shelf with about 100 different values in stock. The only odd ball one was the electrolytic on the board the large 63 Vdc cap. But if i remember correctly LOTS of 10uF.

    • @sebastian_harnisch
      @sebastian_harnisch 7 หลายเดือนก่อน

      @@ZenwizardStudios No worries!

    • @ofgjf
      @ofgjf 7 หลายเดือนก่อน

      You can also exchange the PSU to a newer one. Not all rails of the DELTA PSU are used. Just the 5V, 12V and -12V are needed. Mine is working fine with a meanwell RPT-60B.

    • @sebastian_harnisch
      @sebastian_harnisch 7 หลายเดือนก่อน

      @@ofgjf Thanks for the info! That's interesting.