Excellent explanation. Other points are testing qc material out too long to deteriorate. Uric acid is a dynamic type of analyte. Or the calibrator was made up without leaving big pipette tip in deionized H2O for few seconds and aspiration and dispensing angles. Slow pipetting and steady is key 🔑
Thanks for watching! If you liked the video, please hit the like button, if you want to see more videos, hit the subscribe button, and if you have something to say, please leave a comment.
Excellent explanation. Other points are testing qc material out too long to deteriorate. Uric acid is a dynamic type of analyte. Or the calibrator was made up without leaving big pipette tip in deionized H2O for few seconds and aspiration and dispensing angles. Slow pipetting and steady is key 🔑
Thanks for watching, I have covered some of these points in other video "Calibration in medical Lab". Thanks a lot.
Sir if water is contaminated all results must fail, why only glucose and amylase have failed.?
That example only Glucose and Cholesterol were calibrated before running the QC others were not calibrated. Thanks for watching.