XPS Fitting of Adventitious Carbon

แชร์
ฝัง
  • เผยแพร่เมื่อ 8 ม.ค. 2025

ความคิดเห็น •

  • @albanamarchesini3894
    @albanamarchesini3894 5 ปีที่แล้ว

    thanks! why you put 1.2 to the fwhm. I use to put less than 2. there is some publication where you found that??

  • @vv-ky4bi
    @vv-ky4bi 2 ปีที่แล้ว

    sir, how you create only three curves? if we keep on creating components we get curves....

  • @xavik
    @xavik 7 ปีที่แล้ว

    Thanks. You had to tell the software the number of peaks. So I suppose that you need first to know all the element forms. For example, if a sample has different nitrogen functional groups but you don't know which ones exactly, this wouldn't work because you can't really tell how many peaks it really should have.

  • @chenhaoji9768
    @chenhaoji9768 5 ปีที่แล้ว

    Hi Mark, is XPS able to differentiate C-O-C and C-N content in C 1s?

    • @markbiesinger
      @markbiesinger  5 ปีที่แล้ว +2

      Hi Chenhao, it is tough to do in the C 1s spectrum as there is not a lot of separtation between the C-C, C-N and C-O peaks. I usually infer the amount of C-N in the C 1s spectrum by using information from the N 1s spectrum and the amount of N from the survey spectrum. For example, if I know I only have amine functionality from the N 1s and I know the at. % amount of N present - I can calculate the amount of C-N in the C 1s spectrum and fix the C C-N peak to that amount.

    • @chenhaoji9768
      @chenhaoji9768 5 ปีที่แล้ว

      @@markbiesinger I got it! Thanks very much for your help!

  • @bhawnajoshi8937
    @bhawnajoshi8937 ปีที่แล้ว

    Thanks ❤

  • @urmilabenrathod6762
    @urmilabenrathod6762 7 ปีที่แล้ว

    how you know the position of each peak? Is there any standard values?

    • @markbiesinger
      @markbiesinger  7 ปีที่แล้ว +1

      Yes there are databases for binding energy positions, examples at: www.xpsfitting.com/2008/10/polymer-databases.html

    • @jahjahsun-3
      @jahjahsun-3 4 ปีที่แล้ว

      @@markbiesinger Thank you, very nice video. Do you mean in appendix 1 in Briggs & Beamson's book? I found different shifts in this old paper "Physica Scripta. Vol. 2, 70-80, 1970" and tried to understand more here "Progress in Materials Science 107 (2020) 100591". Where do you suggest to study more of this topic?
      Thanks again.

  • @TengriTR
    @TengriTR 3 ปีที่แล้ว

    thanks

  • @melakbirara2845
    @melakbirara2845 3 ปีที่แล้ว

    it is not visible.