  • @edwinlevis7695
    @edwinlevis7695 4 หลายเดือนก่อน +1

    Hello there Mr. Better BIomed....I'm a student doing a diploma in biomedical engineering.
    This video has really been helpful and informative for this project I am currently working on.
    Be blessed and I hope to see more.

    • @BetterBiomedChannel
      @BetterBiomedChannel  4 หลายเดือนก่อน

      @edwinlevis7695 Thanks for writing! Check out my Biomed Basics playlist and my troubleshooting & repair playlist

  • @zaprodk
    @zaprodk 3 ปีที่แล้ว

    Since the temperature is adjustable on the front panel the mechanical cutout is regulating in case the electronics get unresponsive. There's also a thermal fuse in the heating element, a last resort when all else fails.

  • @zaprodk
    @zaprodk 3 ปีที่แล้ว +2

    The heater is running on AC. In that case it's a TRIAC on the heatsink. The transformer is too small for powering that heating element. I guess there's a small TRIAC for the fan and then the big on the heatsink for the heater.

    • @BetterBiomedChannel
      @BetterBiomedChannel  3 ปีที่แล้ว +1

      I completly agree on the triac for the heater. The fan looks to be relay controlled straight from Mains. I will get better pictures of the power board along with some part numbers. I had 10 minutes to film and prep for this video before other Biomeds come walking through the warehouse.

  • @w8lvradio
    @w8lvradio 3 ปีที่แล้ว

    Ever see a "Bair Hugger" heater? Just wondered... All the Best! 73 DE W8LV BILL