Optiflux 4000 series new model, coil resistance value 120- 180 Ohm, max. 200 Ohm, old model flow sensor 40-80 Ohm 1",1/12",2",3" - 70 Ohm to 120 Ohm , 6",12"14",- Approx 80 to 130 Ohm( may be 10,+,-)
Sir ji Display me 5 wires aa rha h Sensor se 2 coil ka 3 signal Sensor ka Blue color ka jo display meter h waha connect ho rhe h Inme se consa wire disconnect corenge tau meter reading Lena band kr Dena?? Me chhta hu flow on ho tab bhi read ni kare zero dikhae??
@@iandcknow9459 Sir ji Pani ka flow tau h Pani hum usme se pas kr rhe h Leki hum chahte h ke read ni kare flow ko Display me running flow count zero dikhae Lekin Pani ja raha hoga flow ja raha hoga
electrode resistance kitna hai, agr shi mile to convertor side koi problem hai ,electrode wire 2&3 ko shot kr ke dekhiye flow zero aata hai ki nhi, coil current kitna hai,
Thanks , informative vedio
Thanks sir
👌👌👌
Very good
Good
Bhai dhakan kab khologe
Bhai instrumentation k liye interview guidence chahiye thi
Ok, kya jankari chahiye
I have optiflux4000f sensor 14" size.
Please provide resistance value
Optiflux 4000 series new model, coil resistance value 120- 180 Ohm, max. 200 Ohm, old model flow sensor 40-80 Ohm
1",1/12",2",3" - 70 Ohm to 120 Ohm ,
6",12"14",- Approx 80 to 130 Ohm( may be 10,+,-)
Sir flow A+, A per 4-20 a jayega per online Datalogger per totalizer kese dikhega
Ye jo output hai 4-20 mA ye totalizer ke input me connect kro, udher bhi 4-20 mA +&- diya rahta hai
Sir ji
Display me
5 wires aa rha h
Sensor se
2 coil ka
3 signal Sensor ka
Blue color ka jo display meter h waha connect ho rhe h
Inme se consa wire disconnect corenge tau meter reading Lena band kr Dena??
Me chhta hu flow on ho tab bhi read ni kare zero dikhae??
Agr line se flow nhi ja Raha hai aur kuchh value show hi rha hai to ,uske liye empty pipe detection on kr dijiye
Flow zero show karega
Coil wire nikal denge tb bhi zero hoga , sensor wire nikal denge tb zero hoga , aur agr sensor wire ke dono wire ko shot kr denge tb bhi zero hoga
@@iandcknow9459
Sir ji
Pani ka flow tau h
Pani hum usme se pas kr rhe h
Leki hum chahte h ke read ni kare flow ko
Display me running flow count zero dikhae
Lekin Pani ja raha hoga flow ja raha hoga
Wire disconnect kr dijiye, ya jo sensor wire hai 2,3 ko apas me shot kr dijiye
@@iandcknow9459
OK
Thank u 🙏
Line pe flow na hone par v full scale flow dikha raha hai. 20mA aa raha hai. Kaise thik kare?
Aur shield ka termination kaisa hoga.
electrode resistance kitna hai, agr shi mile to convertor side koi problem hai ,electrode wire 2&3 ko shot kr ke dekhiye flow zero aata hai ki nhi, coil current kitna hai,
Dear Sir, How to reset password???
Setup - special function - yha password se related functions mil jayega