How To Align A Scanning Electron Microscope (SEM) | Instructinate

แชร์
ฝัง
  • เผยแพร่เมื่อ 13 ธ.ค. 2024

ความคิดเห็น • 10

  • @il83akafly
    @il83akafly หลายเดือนก่อน

    Great tips for beginners!
    Thanks for sharing!
    Few things for more accurate (optimal) alignments:
    1. better to start from the gun alignment (if it is allowed on your SEM/in your account). ask advanced user how to do on your SEM.
    2. always (!) use Degauss after Wobbling. + check with manual defocusing after degauss if it is still correct.
    3. My personal sequence for the Stigmators: Focus -> Stig A -> Stig B -> Stig A -> Focus.

  • @donnadbarlow
    @donnadbarlow 2 หลายเดือนก่อน

    One of my professors says, "STIGMATION NATION" !!

  • @fijismart1
    @fijismart1 2 ปีที่แล้ว +2

    Nice video and graphics. In your light microscopy analogy you're describing parfocal, but said confocal. Confocal is a light microscopy technique to achieve better resolution, contrast, and depth of field.

    • @instructinate
      @instructinate  2 ปีที่แล้ว

      You're right, I said confocal instead of PARFOCAL. oops.

  • @thaisparadella3159
    @thaisparadella3159 7 หลายเดือนก่อน

    What an amazing video!!! I really enjoyed! I have one question, though. In some FEI Inspect SEM (I work with a tungsten Inspect S50 from FEI SEM), there isn't the Wobbler button, or Wobbler option. No problems in adjusting the stigmators, however, without the Wobbler, I just turn on the "autofocus" and proceed with the rest that you have shown in the video? Thanks in advance!

  • @tonyduong70
    @tonyduong70 ปีที่แล้ว

    thank you !

  • @maxmustermann850
    @maxmustermann850 ปีที่แล้ว

    Great video! I have a few question about the SEM:
    1. what is the working distance? Is this the focus?
    2. what is the difference between SEM magnification and print magnification?

    • @instructinate
      @instructinate  ปีที่แล้ว

      WD can be a lot of things to a lot of people. My understanding is it is the distance from the pole piece to where the beam comes at crossover (focus).
      There can be a huge difference between screen mag and print mag. many newer SEM's will give you the option of which to use. Since mag is so squishy, the standard today is to use pixel size or the horizontal width of the image.

  • @triggerfingerstudios
    @triggerfingerstudios 2 ปีที่แล้ว

    Stigmators 4 life!

  • @jiggysiggy
    @jiggysiggy 4 หลายเดือนก่อน

    G