Calibrate Theta2Theta on Bruker D8 Advance using D8 Tools

แชร์
ฝัง
  • เผยแพร่เมื่อ 29 ก.ย. 2024
  • Sorry the video cuts short. On that final page, move the drive slightly down and then find reference up. Do for both theta and 2 theta.
    Fix for when Bruker D8 Advance RAM is bad and has been power cycled so theta and 2 theta positions are unknown to the software. May need to double check the found positions are accurate by moving the theta/2theta to 30 and 60 which should result in a horizontal sample stage and symmetrical detector to the source (x-ray tube)

ความคิดเห็น • 2

  • @МишаПетров-ю6б
    @МишаПетров-ю6б 3 ปีที่แล้ว