- Sorry the video cuts short. On that final page, move the drive slightly down and then find reference up. Do for both theta and 2 theta.
Fix for when Bruker D8 Advance RAM is bad and has been power cycled so theta and 2 theta positions are unknown to the software. May need to double check the found positions are accurate by moving the theta/2theta to 30 and 60 which should result in a horizontal sample stage and symmetrical detector to the source (x-ray tube)
Hello, please tell me if it is possible to contact you by email? I want to ask you a question about installing Bruker.
can't hear anything :-(