Multiple Skytron Surgical Table Problems

แชร์
ฝัง
  • เผยแพร่เมื่อ 26 พ.ย. 2024

ความคิดเห็น • 14

  • @christianbond8299
    @christianbond8299 3 ปีที่แล้ว +3

    Great explanations for so many key aspects of Logic and Troubleshooting the BMET way. Keep these types of videos coming our way!

  • @willb914
    @willb914 2 ปีที่แล้ว

    Excellant pointers and solutions. Thanks Much !!!

  • @acmedepo4286
    @acmedepo4286 2 ปีที่แล้ว +2

    Thank you, you're insight/explanation was on the money, blown mini fuse. .Appreciate the help.

  • @takingrainbowsback
    @takingrainbowsback 2 ปีที่แล้ว +2

    Im having, I think a similar issue where the back wont go all the way up.. the kidney rest is down and will not go up or even make a motor sound like it's trying too... it will make a motor sound when I hit the kidney down button. But it's the Old 6500 hd with the red sticker.

    • @BetterBiomedChannel
      @BetterBiomedChannel  2 ปีที่แล้ว +1

      yes, there are collision microswitches and one of them is out of adjustment.

  • @MrBiomedical
    @MrBiomedical 3 ปีที่แล้ว +1

    Good explanation♥️

  • @ouahchiamor7627
    @ouahchiamor7627 ปีที่แล้ว

    Hello,can you tell me the periodic maintenance to change oil and battery

  • @haissam326500
    @haissam326500 ปีที่แล้ว

    Do you have the part number or ratings for the SOC 1.6 A fuse. And voltage thank you

  • @BjornV78
    @BjornV78 3 ปีที่แล้ว

    With the microswitch problem, why didn't they use a small hole in the moving rod of the kidney bridge, and a lightsensor/detector ?
    That way there are less mechanical parts and less wear and tear.

    • @BetterBiomedChannel
      @BetterBiomedChannel  3 ปีที่แล้ว +1

      The microswitch is cheap, sealed, easy to adjust and test compared to an optical detector. An optical detector could give a false positive if blood or debris fell down there - which is guaranteed since it's surgery.

  • @quensyatzabita59
    @quensyatzabita59 2 ปีที่แล้ว

    completion of the Schaerer AXIS 400 operating table, error code E12: Calibrating Data Failure. Data from position sensors doesn't match up with main controller board's data. The table must be recalibrated.
    please help master Better Biomed

    • @BetterBiomedChannel
      @BetterBiomedChannel  2 ปีที่แล้ว

      I have never heard of that table. I will do some research.

    • @quensyatzabita59
      @quensyatzabita59 2 ปีที่แล้ว

      please help sir. on the remote it shows E12.